Digital Differential Pressure Sensor with 4-20 Ma Hart Protocol
-
Payment
-
Origin
China Mainland
-
Minimum Order
1
-
Packing
Pieces
- Contact Now Start Order
- Description
Product Detail
Digital Differential Pressure Sensor with 4-20 Ma Hart Protocol
Production Introduction:
Model:QP-86D differential pressure sensor
Dimensions :( unit: mm)
Span:
Range code |
Range |
Measure span |
|
MIN |
MAX |
||
4 |
0.374KPa |
37.4KPa |
-37.4~37.4KPa |
5 |
1.868KPa |
186.8KPa |
-186.8~186.8KPa |
6 |
6.9KPa |
690KPa |
-690~609KPa |
7 |
20.68KPa |
2068KPa |
-690~609KPa |
Output:
Analog output: two wire 4~20mADC
Digital output: two wire 4~20mADC and variable digital signal in accordance with HART AGREEMENT.
Power: supply voltage 12.5~45VDC
Limit load: Max load resistor Rmax in loop decided by the supply voltage, see the below drawing. HART AGREEMENY require the load resistor not less than 250 O, not more than 600O
Communication distance: Max. 1.5km, multi-address communication can connect 15 smart transmitters max.
Output display: 4 digital display, 5 digit LCD display, pointer type.
Marking: ExiaII CT5; explosion-proof isolation type
Zero offset: can positive offset or negative offset, but can not overstep the utmost value
Temperature: environment temperature-40~85? (When choose LCD:-20~70?)
Medium temperature:-40~104?
Static pressure and one side overload: 3.45Kpa(Absolute) 31.2MPa(Gauge)
Relative humidity:5 ~98%RH
Startup time: less than 5 sec.
Capability change: less than 4px3
Damp adjust: 0.20~25sec.
Malfunction alarming: If appear malfunction, automatism diagnose function will out 3.9 or 21.0mA
Protection: To avoid the datas are modified random, can utilize
Software or protect switch lock the memory data
Reference condition: zero is the lower limit and not offset, temperature 25?, Atmospheric pressure, 24VDC power, silicon oil, diaphragm 316L, data adjust range the same as the up and lower limit of the range
Accuracy: ±0.1%, adjust range=0.1URL
±[0.05+(0.005URL/span)]%, adjust range 0.025URL~0.1UPL
±[0.01+(0.006URL/span)]%, adjust range 0.01URL~0.025URLNote: accuracy have include the influence of linearlity, hysteresis and repeatability
Stability: ±0.1%URL/year
Tem.influence: ±(0.05%URL+0.15%span)/20?
Static pressure influence: ±zero error 0.1URL/6.9Mpa
Libration effect: withstand frequency 200Hz at any direction, arouse error ±0.05%URL
Power influence: 0.005%/V
Load influence: no load influence.
Install position influence: about 250Pa zero offset, can be eliminate via calibration, no range influence.
RF interference: drift is ±0.1%span in the condition of 30-1000MHz frequency and the most field intensity 10V/m
- Aluminum Platform Scale Weighing Load Cells 100kg 500kg 1000kg 2000kg 2500kg 1 Pieces / (Min. Order)
- 15kg To 100kg Alloy Aluminum Single Point Weight Load Cell 1 Pieces / (Min. Order)
- 100kg To 500kg Single Point Weight Load Cell 1 Pieces / (Min. Order)
- Digital Oil Pressure Gauge with Progress Bar and Battery Power Pressure Gauge 1 Pieces / (Min. Order)
- 50kg Small Load Cell 1 Pieces / (Min. Order)
- High Precision Parallel Beam Load Cell 1 Pieces / (Min. Order)
- Electronic Scale Load Cell with High Precision 1 Pieces / (Min. Order)
- CE,OIML Certificate Platform Scale Load Cell Sensor for Weighing System 1 Pieces / (Min. Order)
- Single Point Load Cell for Weighing Scale 1 Pieces / (Min. Order)
- Aluminum Alloy Strain Gauge Load Cell 100kg 1 Pieces / (Min. Order)